Zeile 17: | Zeile 17: | ||
For both problems I gladly found the solution. The approach is like this: | For both problems I gladly found the solution. The approach is like this: | ||
* set the sampling counter to MAX (600 Mio) | * set the sampling counter to MAX (600 Mio) | ||
- | * clear the buffer with '' | + | * clear the buffer with '' |
* set trigger source to ' | * set trigger source to ' | ||
* initiate the DMM (arms the trigger) | * initiate the DMM (arms the trigger) | ||
Zeile 23: | Zeile 23: | ||
* read samples with '' | * read samples with '' | ||
- | + | When the DMM isn't asked for any value, it will sample with the preset NPLC setting (200ms by default) and therefore also updates the LCD reading. If we want the latest value from the memory, we query for it with '' | |
===== Python Code ===== | ===== Python Code ===== | ||
Below you can find a python class for communicating with it. It implements the upper approach. | Below you can find a python class for communicating with it. It implements the upper approach. | ||
Zeile 230: | Zeile 229: | ||
===== Links ===== | ===== Links ===== | ||
[1] https:// | [1] https:// | ||
- | [2] https:// | + | [2] https:// |
[3] https:// | [3] https:// | ||